Julian Lommen,Lara Schorn,Christoph Sproll,Norbert R. Kübler,Luis Fernando Nicolini,Ricarda Merfort,Ayimire Dilimulati,Frank Hildebrand,Majeed Rana,Johannes Greven
摘要
Mandible defects are conventionally reconstructed using titanium plates. However, titanium causes metallic artifacts which impair radiological imaging. This study aims at evaluating mechanical fatigue of radiolucent fiber-reinforced polyetheretherketone (f-PEEK), polyetheretherketone (PEEK), polyetherketoneketone (PEKK), and polyphenylsulfone (PPSU) polymer plates for mandible reconstruction. A total of 30 plates (titanium [n = 6], f-PEEK [n = 6], PEEK [n = 6], PEKK [n = 6], PPSU [n = 6]) were implanted in synthetic mandibulectomized polyurethane mandibles. Servo-pneumatic mechanical testing with cyclic application of 30–300 N at 3 Hz was conducted. Bite forces were 70% on the unresected and 30% on the resected side. Total number of cycles was set to 250,000. Testing was aborted in case of plate or screw failure. Axial load to failure was tested with a speed of 1 mm/s. Kruskal–Wallis and Dunn’s post hoc tests were used. Titanium, f-PEEK, and PEEK showed no failure in fatigue testing and PPSU (p < 0.001) failed against titanium, f-PEEK, PEEK, and PEKK. Titanium allowed the highest load to failure compared to f-PEEK (p = 0.049), PEEK (p = 0.008), PEKK (p < 0.001), and PPSU (p = 0.007). f-PEEK, PEEK, and PEKK withstood expected physiological bite force. Although titanium plates provided the highest fatigue strength, f-PEEK and PEEK plates showed no failure over 250,000 chewing cycles indicating sufficient mechanical strength for mandible reconstruction.