医学
血管闭合器
并发症
结束语(心理学)
外科
假性动脉瘤
纤维接头
动脉瘤
栓塞
放射科
股动脉
市场经济
经济
作者
Larab L Giniyani,Yesha Rana,Balaram Krishna J Hanumanthu,Doris Chan,Tak Kwan
出处
期刊:Future Cardiology
[Future Medicine]
日期:2021-10-01
卷期号:17 (7): 1193-1197
被引量:4
标识
DOI:10.2217/fca-2020-0154
摘要
Background: Vascular closure devices have replaced mechanical compression for closure of femoral access sites after endovascular procedures. Case: We present an 87-year-old male with symptomatic infrarenal aortic aneurysm measuring 4.8 cm presenting for elective endovascular repair of the aortic aneurysm. A Perclose ProGlide Suture-Mediated Closure was used for closure. The closure was complicated by a separation of the ProGlide device resulting in the migration of the footplate to the descending aorta. Correction required snare retrieval via radial access, and the patient recovered without complications. Discussion: We highlight an important complication of the Perclose ProGlide Suture-Mediated Closure device that is rare but important to know when performing endovascular closures with this device.
科研通智能强力驱动
Strongly Powered by AbleSci AI