Vickram Vignesh Rangaswamy,Shunmuga Sundaram Ponnusamy
出处
期刊:Heart Rhythm
[Elsevier]
日期:2024-04-01
卷期号:21 (4): 490-491
标识
DOI:10.1016/j.hrthm.2023.12.023
摘要
A 75-year-old man with intermittent atrioventricular block and left ventricular dysfunction underwent successful left bundle branch area pacing after 4 attempts using the C315HIS catheter and 3830 SelectSecure lumenless lead (Medtronic, Minneapolis, MN) by premature ventricular complex–guided approach1 (Figure 1A; Online Supplemental Video 1). The unipolar pacing threshold was 0.8 V/0.5 ms with a sensed R-wave amplitude of 7 mV and a unipolar pacing impedance of 640 Ω. The delivery catheter was removed after providing lead slack in the atrium (Figure 1A).