Ultrasensitive and Reusable Graphene Oxide-Modified Double-Interdigitated Capacitive (DIDC) Sensing Chip for Detecting SARS-CoV-2.
纳米技术
氧化物
作者
Parshant Kumar Sharma,Eun Seong Kim,Sachin Mishra,Enkhzaya Ganbold,Ryun-Sang Seong,Ajeet Kaushik,Nam-Young Kim
出处
期刊:ACS Sensors [American Chemical Society] 日期:2021-09-24卷期号:6 (9): 3468-3476被引量:10
标识
DOI:10.1021/acssensors.1c01437
摘要
This research reveals the promising functionalization of graphene oxide (GrO)-glazed double-interdigitated capacitive (DIDC) biosensing platform to detect severe acute respiratory syndrome coronavirus (SARS-CoV-2) spike (S1) proteins with enhanced selectivity and rapid response. The DIDC bioactive surface consisting of Pt/Ti featured SiO2 substrate was fabricated using GrO/EDC-NHS/anti-SARS-CoV-2 antibodies (Abs) which is having layer-by-layer interface self-assembly chemistry method. This electroactive immune-sensing platform exhibits reproducibility and sensitivity with reference to the S1 protein of SARS-CoV-2. The outcomes of analytical studies confirm that GrO provided a desired engineered surface for Abs immobilization and amplified capacitance to achieve a wide detection range (1.0 mg/mL to 1.0 fg/mL), low limit of detection (1 fg/mL) within 3 s of response time, good linearity (18.56 nF/g), and a high sensitivity of 1.0 fg/mL. Importantly, the unique biochip was selective against blood-borne antigens and standby for 10 days at 5 °C. Our developed DIDC-based SARS-CoV-2 biosensor is suitable for point-of-care (POC) diagnostic applications due to portability and scaling-up ability. In addition, this sensing platform can be modified for the early diagnosis of severe viral infections using real samples.