Canwei Mao,Ke Zhao,Cyrus Foo Jit Loong,Dean Song,Grzegorz Lisak
出处
期刊:ACS Sensors
[American Chemical Society]
日期:2024-08-08
卷期号:9 (8): 3870-3876
被引量:1
标识
DOI:10.1021/acssensors.4c01673
摘要
Ion-selective electrodes (ISEs) have widespread use in the fields of clinical and environmental analyses. Tetrahydrofuran (THF) is the most used solvent for the preparation of modern ISEs, equipped with ion-selective membranes (ISMs). Until now, the influence of impurities in THF toward potentiometric instability of ion-selective membrane based ISEs was probably associated with the presence of either residual water or peroxide. To address this issue, most literature recommends redistilling THF prior to use in the preparation of the potentiometric membranes. Current study reveals that the actual THF impurity that is responsible for potential instability in the ISM includes products from the oxidation of THF, which contains the hydroxyl group and possibly carbonyl group with a boiling point of above 200 °C. The density functional theory calculation supported pathway of the chemical reaction of THF oxidation, hence, the chemical structure of the uncertain impurities was predicted. The underlying reason for the deteriorating potential stability of the ISEs is proposed as the significant hydrophilicity of these impurities that affect the partitioning of the ion sensing components in the membrane, thus enhancing the leaching of the membrane components from the membrane phase. This finding explains why redistillation of aged THF is advised.