Mengyang Huang,Fang Peng,Shixue Guan,Jiawei Zhang,Wenjia Liang,Junpu Wang,Yi Tian,Yuzhu Su,Yue Tang,Wenqiang Wang,Duanwei He
摘要
Pressure is the core of high-pressure science and technology, and the accuracy of pressure calibration is of much importance for high-pressure experiments and production. Although the pressure limit of the large volume press (LVP) continues to increase, there are no well solutions for in situ pressure calibration. In this study, using in situ high-pressure electrical performance measurement technology, two ideal calibration standard materials in powder conductors, cadmium phosphide (Cd3P2) and zinc telluride (ZnTe) with stable physical and chemical properties and obvious resistance change, are applied to pressure calibration in the LVP. In situ high-pressure synchrotron radiation x-ray diffraction was used to verify the phase transition pressure point of Cd3P2. The introduction of powder conductors for pressure calibration commits to establish a pressure system, which is safer, more stable to operate, and more accurate in experimental measurements for the LVP.