Naren Shetty,Luci Kaweri,Ann Sarah Koshy,Rohit Shetty,Rudy M.M.A. Nuijts,Abhijit Sinha Roy
标识
DOI:10.1097/j.jcrs.0000000000000494
摘要
Purpose: To compare the repeatability of ocular biometry measured with the LENSTAR LS 900, IOLMaster 700, and Anterion and its impact on predicted intraocular lens (IOL) power. Setting: Tertiary eye-care facility in South India. Design: Prospective, observational, cross-sectional study. Methods: Eyes diagnosed with cataract had 3 consecutive scans on each biometers. The repeatability was assessed using the within-subject standard deviation (S w ), test-retest repeatability, and coefficient of variation (CoV). The agreement was evaluated with the intraclass correlation (ICC). The IOL power was calculated with the Barrett Universal II formula. Results: The study comprised 127 eyes of 76 patients. The repeatability of all parameters for a given device were excellent (ICC >0.9, low CoV and S w ). The agreement of the parameters between the biometers was very good (range from 0.93 to 0.99). The predicted IOL power differed statistically between the devices ( P < .05), but the difference was clinically insignificant between the 3 biometers (ICC >0.99 for repeat calculation of IOL power). Conclusions: All the biometers included in the study had good to excellent repeatability of biometry parameters. The agreement of the predicted IOL power was excellent between the 3 optical biometers.