二尖瓣环
对偶(语法数字)
内科学
心脏病学
环空(植物学)
二尖瓣
医学
材料科学
复合材料
艺术
血压
舒张期
文学类
标识
DOI:10.20944/preprints202506.0778.v1
摘要
Background: The development of transcatheter mitral valves (TMVs) represents a major advancement in cardiology, driven in part by the growing elderly population. These patients frequently suffer from secondary mitral regurgitation but are often ineligible for surgical valve replacement due to high procedural risks. This study aimed to develop a self-expanding TMV stent fabricated from a single nitinol tube, featuring two distinct central zones: a smaller-diameter valve-containing segment and a larger-diameter anchoring segment for the mitral annulus. Methods: Biotechnical engineering using COMSOL Multiphysics software package, manufacturing prototypes of stents and valves in 5 sizes, 22 Fr delivery system compatibility assessment, and pulsatile-flow testing. Results: We bioengineered a novel stent design with an integrated porcine pericardial valve. The stents were laser-cut from nitinol tubes (4.5 mm outer diameter, 0.45 mm wall thickness) and heat-treated to achieve spatial configurations compatible with fibrous ring diameters of 40, 42, 44, 46, and 48 mm. Pericardial leaflets and coverings were then mounted onto the stents. The resulting valves were successfully loaded into a 24 Fr delivery system and exhibited proper opening and closing function under pulsatile-flow testing. Conclusions: Our findings confirm the feasibility of a single-component, dual-diameter TMV stent, offering a promising solution for high-risk patients with mitral regurgitation.