<b><i>Background:</i></b> The mode of waveform generation and circuit characteristics differ between high-frequency oscillators. It is unknown if this influences performance. <b><i>Objectives:</i></b> To describe the relationships between set and delivered pressure amplitude (&#x0394;P), and the interaction with frequency and endotracheal tube (ETT) diameter, in eight high-frequency oscillators. <b><i>Methods:</i></b> Oscillators were evaluated using a 70-ml test lung at 1.0 and 2.0 ml/cm H<sub>2</sub>O compliance, with mean airway pressures (P<sub>AW</sub>) of 10 and 20 cm H<sub>2</sub>O, frequencies of 5, 10 and 15 Hz, and an ETT diameter of 2.5 and 3.5 mm. At each permutation of P<sub>AW</sub>, frequency and ETT, the set &#x0394;P was sequentially increased from 15 to 50 cm H<sub>2</sub>O, or from 20 to 100% maximum amplitude (10% increments) depending on the oscillator design. The &#x0394;P at the ventilator (&#x0394;P<sub>VENT</sub>), airway opening (&#x0394;P<sub>AO</sub>) and within the test lung (&#x0394;P<sub>TRACH</sub>), and tidal volume (V<sub>T</sub>) at the airway opening were determined at each set &#x0394;P. <b><i>Results:</i></b> In two oscillators the relationships between set and delivered &#x0394;P were non-linear, with a plateau in &#x0394;P thresholds noted at all frequencies (Dräger Babylog 8000) or ≥10 Hz (Dräger VN500). In all other devices there was a linear relationship between &#x0394;P<sub>VENT</sub>, &#x0394;P<sub>AO</sub> and &#x0394;P<sub>TRACH</sub> (all r<sup>2</sup> >0.93), with differing attenuation of the pressure wave. Delivered V<sub>T</sub> at the different settings tested varied between devices, with some unable to deliver V<sub>T</sub> >3 ml at 15 Hz, and others generating V<sub>T</sub> >20 ml at 5 Hz and a 1:1 inspiratory-to-expiratory time ratio. <b><i>Conclusions:</i></b> Clinicians should be aware that modern high-frequency oscillators exhibit important differences in the delivered &#x0394;P and V<sub>T</sub>.

科研通智能强力驱动
Strongly Powered by AbleSci AI
科研通是完全免费的文献互助平台,具备全网最快的应助速度,最高的求助完成率。 对每一个文献求助,科研通都将尽心尽力,给求助人一个满意的交代。
实时播报
爆米花应助veronica采纳,获得10
1秒前
爆米花应助Amberwdd采纳,获得10
1秒前
111关闭了111文献求助
2秒前
2秒前
7秒前
大气的草莓完成签到,获得积分10
7秒前
wuhan发布了新的文献求助10
7秒前
Losemisery发布了新的文献求助10
7秒前
8秒前
搜集达人应助孑孑采纳,获得10
9秒前
10秒前
遂安完成签到,获得积分10
11秒前
无情寻芹发布了新的文献求助20
13秒前
Joyhold完成签到,获得积分10
14秒前
甜甜凌青发布了新的文献求助10
15秒前
云为翳完成签到,获得积分10
17秒前
18秒前
cong完成签到 ,获得积分10
18秒前
George完成签到,获得积分10
19秒前
北冥有鱼完成签到 ,获得积分10
19秒前
22秒前
研友_nEoBP8完成签到,获得积分10
23秒前
简单山水发布了新的文献求助10
24秒前
27秒前
27秒前
28秒前
量子星尘发布了新的文献求助10
29秒前
亮之完成签到,获得积分10
29秒前
29秒前
29秒前
wuqs发布了新的文献求助10
33秒前
33秒前
是否发布了新的文献求助10
33秒前
34秒前
35秒前
35秒前
genomed完成签到,获得积分0
35秒前
凉凉完成签到,获得积分10
38秒前
39秒前
39秒前
高分求助中
(应助此贴封号)【重要!!请各用户(尤其是新用户)详细阅读】【科研通的精品贴汇总】 10000
Burger's Medicinal Chemistry, Drug Discovery and Development, Volumes 1 - 8, 8 Volume Set, 8th Edition 1800
Cronologia da história de Macau 1600
Contemporary Debates in Epistemology (3rd Edition) 1000
International Arbitration Law and Practice 1000
文献PREDICTION EQUATIONS FOR SHIPS' TURNING CIRCLES或期刊Transactions of the North East Coast Institution of Engineers and Shipbuilders第95卷 1000
BRITTLE FRACTURE IN WELDED SHIPS 1000
热门求助领域 (近24小时)
化学 材料科学 医学 生物 工程类 有机化学 纳米技术 计算机科学 化学工程 生物化学 物理 复合材料 内科学 催化作用 物理化学 光电子学 细胞生物学 基因 电极 遗传学
热门帖子
关注 科研通微信公众号,转发送积分 6160507
求助须知:如何正确求助?哪些是违规求助? 7988803
关于积分的说明 16605888
捐赠科研通 5268738
什么是DOI,文献DOI怎么找? 2811185
邀请新用户注册赠送积分活动 1791287
关于科研通互助平台的介绍 1658155