角膜曲率计
激光矫视
人工晶状体
医学
角膜地形图
眼科
白内障手术
视力
验光服务
住宿
角膜
光学
物理
作者
Edoardo Ligabue,C. Giordano
标识
DOI:10.3928/1081597x-20090115-05
摘要
Purpose: To present the use of the point spread function (PSF) metric pre- and postoperatively for the assessment of visual quality in cataract and refractive surgery. Methods: Case examples of cataract and refractive surgery and the effect of accommodation are presented. All PSF measurements were obtained using the NIDEK OPD-Scan II, and corneal internal and total aberrations were simulated using the NIDEK OPD-Station software. All eyes underwent corneal topography, wavefront aberrometry, autorefraction, keratometry, and pupillometry measurements pre- and postoperatively using the OPDScan II. The PSF was used to assess visual quality. Results: Four case examples including refractive surgery, aspheric multifocal intraocular lens (IOL) implantation, toric IOL implantation, and the difference in PSF due to accommodation are presented. Conclusions: The PSF in refractive and cataract surgery is a clinically valuable metric to evaluate the patient's visual quality pre- and postoperatively. The OPD-Scan II can measure the effect of accommodation and distant fixation providing an assessment of changes in visual quality. [ J Refract Surg . 2009;25:S104–S109.]