Design of Interrogation Protocols for Radiation Dose Measurements Using Optically-Stimulated Luminescent Dosimeters
作者
Sara A. Abraham,Kimberlee J. Kearfott,Ali H. Jawad,Andrew J. Boria,Tobias J. Buth,Alexander Dawson,Sheldon C. Eng,Samuel J. Frank,Crystal A. Green,Mitchell L. Jacobs,Kevin Liu,Joseph A. Miklos,Hien Van Nguyen,Muhammad Rafique,Blake D. Rucinski,T. J. Smith,Yan-Liang Tan
出处
期刊:Health Physics [Lippincott Williams & Wilkins] 日期:2017-01-25卷期号:112 (3): 237-245被引量:5
Optically-stimulated luminescent dosimeters are capable of being interrogated multiple times post-irradiation. Each interrogation removes a fraction of the signal stored within the optically-stimulated luminescent dosimeter. This signal loss must be corrected to avoid systematic errors in estimating the average signal of a series of optically-stimulated luminescent dosimeter interrogations and requires a minimum number of consecutive readings to determine an average signal that is within a desired accuracy of the true signal with a desired statistical confidence. This paper establishes a technical basis for determining the required number of readings for a particular application of these dosimeters when using certain OSL dosimetry systems.