Kazuhiro Shimamoto,O Kaii,Mitsunori Ikeda,Yoshihito Takeuchi,Takashi Ishigaki,Shinya Sakuma
出处
期刊:PubMed
日期:1991-12-25
卷期号:51 (12): 1425-30
被引量:1
摘要
Mechanical arc scanning is widely used for breast sonography in Japan. The authors have used three different kinds of devices over the past seven years. The diagnostic accuracy was compared among three groups of patients with pathologically confirmed breast masses. In group A, 309 cases (77 carcinomas) were evaluated with a 10 MHz contact compound scanner, and the sensitivity, specificity and accuracy rates were 77.9%, 97.8% and 92.9%, respectively. In group B, 306 cases (56 carcinomas) were evaluated with a 5 or 7.5 MHz mechanical arc scanner, and the sensitivity, specificity and accuracy rates were 89.3%, 84.8% and 85.6%, respectively. In group C, 296 cases (71 carcinomas) were evaluated with a 7.5 MHz real time scanner, and the sensitivity, specificity and accuracy rates were 95.8%, 92.4% and 93.2%, respectively. In cases with T1 breast cancer, the methods had sensitivities of 60.0%, 85.7% and 94.5%, respectively. The sensitivity of the real-time scanner was not significantly different from that of the mechanical arc scanner. In conclusion, the hand-held real-time scanner with high-frequency transducer is a simple, useful device with high diagnostic accuracy for breast examination and can be used as a substitute for the mechanical arc scanner.