电穿孔
电极阵列
电极
相容性(地球化学)
视网膜
视网膜
转染
材料科学
生物医学工程
纳米技术
化学
生物
工程类
复合材料
生物化学
基因
神经科学
物理化学
作者
Yu Zhang,Tao Peng,Yu Ge,Mengda Li,Chendi Li,Jiyu Xi,Zixi Li,Zewen Wei,Yuntao Hu
出处
期刊:Lab on a Chip
[Royal Society of Chemistry]
日期:2024-01-01
卷期号:24 (7): 1957-1964
被引量:3
摘要
, relatively high electroporation voltage, poor surgical operability and a lack of spatial selectivity of retina tissue) have prevented electroporation from being approved for clinical application (or even clinical trials). In this study, a flexible micro-electrode array for retina electroporation (FERE) was developed for retina electroporation. A suitably shaped flexible substrate and well-placed micro-electrodes were designed to adapt to the retina curvature and generate an evenly distributed electric field on the retina with a significantly reduced electroporation voltage of 5 V. The FERE provided (for the first time) a capability of controlled gene delivery to the different structural layers of retina tissue by precise control of the distribution of the electrical field. After ensuring the surgical operability of the FERE on rabbit eyeballs, the FERE was verified to be capable of transfecting different layers of retina tissue with satisfactory efficiency and minimum damage. Our method bridges the technical gap between laboratory validation and clinical use of retina electroporation.
科研通智能强力驱动
Strongly Powered by AbleSci AI