Aaron R. Todd,Andrew Alexander,Martin F. Jarrold
标识
DOI:10.1021/jasms.9b00010
摘要
Charge detection mass spectrometry (CDMS) depends on the measurement of the charge induced on a cylinder by individual ions by means of a charge-sensitive amplifier. Electrical noise limits the accuracy of the charge measurement and the smallest charge that can be detected. Thermal noise in the feedback resistor is a major source of electrical noise. We describe the implementation of a charge-sensitive amplifier without a feedback resistor. The design has significantly reduced 1/f noise facilitating the detection of high m/z ions and substantially reducing the measurement time required to achieve almost perfect charge accuracy. With the new design we have been able to detect individual ions carrying a single charge. This is an important milestone in the development of CDMS.