Abstract In this work, sensors based quartz crystal microbalance (QCM) technology were functionalized with hexamethyldisiloxane (HMDSO) coatings deposited via plasma-enhanced chemical vapor deposition (PECVD) to enhance their detection of volatile organic compounds (VOCs). The influence of deposition conditions on film morphology, surface chemistry, and sensing performance under both ambient and humid environments was systematically investigated. Higher deposition pressures promoted aggregate formation, increasing surface roughness to 83.7 nm and providing more active sites for VOC adsorption. water contact angles analysis confirmed a transition from hydrophilic to hydrophobic surface properties, while EDS revealed a biphasic structure consisting of a dense silica matrix interspersed with organic-rich clusters generated under high-pressure conditions. These structural characteristics resulted in a twentyfold increase in sensitivity when the monomer partial pressure was elevated, achieving a LOD/LOQ of (2/6.5 ppm) with rapid response/recovery times (120/72 s) and strong linearity (R2 ≥ 0.912). Notably, low-pressure sensors experienced a reduction in ethanol sensitivity under humid conditions due to water interference, whereas high-pressure sensors retained their hydrophobicity, effectively mitigating humidity effects and ensuring stable, reliable VOC detection.