Stefan Bartenschlager,Alexander Cavallaro,Tobias Pogarell,Michael Uder,Oliver Chaudry,Georg Schett,Klaus Engelke
标识
DOI:10.1016/j.ejrad.2025.111923
摘要
OBJECTIVES: Contrast agents are frequently administered in computed tomography (CT) scans used for opportunistic screening of osteoporosis. The objective of this study is to compare the impact of contrast-related bone mineral density (BMD) increase between phantom-based and internal CT calibration techniques. MATERIALS AND METHODS: Phantom-based and internal CT calibration techniques were used to determine trabecular BMD in 93 existing clinical CT scans of the lumbar spine of 34 subjects, scanned before and after administration of contrast agents. BMD differences between native (Nat) and contrast-enhanced scans in arterial (Art) and portal venous (PV) phases were compared among calibration methods. Three pairs of internal reference materials were investigated: blood/air, subcutaneous adipose tissue (SAT)/air and muscle tissue/air. RESULTS: in PV phase) did not significantly differ (p > 0.1 for Art and PV phases) from increases of phantom-calibrated BMD. CONCLUSIONS: Although CT values of internal reference materials increased following the administration of contrast, internal calibration was unable to mitigate the contrast-related BMD increases observed in phantom-based calibration.