Tanyanan Tanawuttiwat,Mithilesh K. Das,John M. Miller,Maya Guglin
出处
期刊:Heart Rhythm
[Elsevier BV]
日期:2023-03-14
卷期号:20 (6): 918-926
被引量:11
标识
DOI:10.1016/j.hrthm.2023.02.020
摘要
The current design of an innovative left ventricular assist device (LVAD) makes use of magnetic levitation technology, which enables the rotors of the device to be completely suspended by magnetic force, reducing friction and blood or plasma damage. However, this electromagnetic field can result in electromagnetic interference (EMI), which can interfere with proper functioning of another cardiac implantable electronic device (CIED) in its direct proximity. Approximately 80% of patients with an LVAD have a CIED, most frequently an implantable cardioverter-defibrillator (ICD). Several device-device interactions have been reported, including EMI-induced inappropriate shocks, inability to establish telemetry connection, EMI-induced premature battery depletion, undersensing by the device, and other CIED malfunctions. Unfortunately, additional procedures, including generator exchange, lead adjustment, and system extraction, are frequently required because of these interactions. In some circumstances, the additional procedure might be preventable or avoidable with appropriate solutions. In this article, we describe how EMI from the LVAD impacts the functionality of the CIED and provide possible management options, including manufacturer-specific information, for the current CIEDs (eg, transvenous and leadless pacemakers, transvenous and subcutaneous ICDs, and transvenous cardiac resynchronization therapy pacemakers and ICDs).
科研通智能强力驱动
Strongly Powered by AbleSci AI