周长
标准差
呼吸
医学
心脏病学
通风(建筑)
生物医学工程
内科学
物理医学与康复
数学
统计
解剖
工程类
机械工程
几何学
作者
Bryson Padasdao,Ehsaneh Shahhaidar,Olga Borić-Lubecke
标识
DOI:10.1109/embc.2013.6609906
摘要
In this paper, an off-the-shelf DC motor is modified into a chest belt and used to successfully measure circumference change on a mechanical chest model, while simultaneously harvesting significant power. Chest circumference change can provide information on tidal volume, which is vital in assessing lung function. The chest circumference change is calculated from the motor's voltage output. Calculated values are within 0.95mm of measured circumference changes, with a standard deviation of 0.37mm. The wearable motor can also harvest at least 29.4µW during normal breathing.