Main text As part of the ongoing BIPM key comparison BIPM.EM-K10.b, a direct comparison of the Josephson array voltage standards of the Bureau International des Poids et Mesures (BIPM) and the National Metrology Institute of Finland (VTT MIKES), Espoo, Finland, was carried out in October 2019 at the level of 10 V. For this exercise, the option A of the BIPM.EM-K10.b comparison protocol was partially applied. Option A required the MIKES to provide a reference voltage with its Josephson voltage standard for measurement by the BIPM using an analogue nanovoltmeter and associated measurement cables. Since no sufficiently stable voltage could be achieved in this configuration, a digital detector was used instead. In all cases the BIPM array was kept floating from ground. The final results were in good agreement within the combined relative standard uncertainty of 2.5 parts in 10 10 for the nominal voltage of 10 V. To reach the main text of this paper, click on Final Report . Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/ . The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).