BACKGROUND: auto-injector is a novel prefilled, single-use, needle-free device currently in development for intradermal, subcutaneous, and intramuscular agent delivery across a variety of clinical indications. We aimed to evaluate delivery depth of the device calibrated at pressure appropriate for intramuscular (IM) administration. METHODS: This was a prospective single-center study in healthy adult volunteers, in whom each received a single injection of saline into the anterolateral thigh. Using sequential MRI scans, we measured skin-to-muscle distance (STMD) agent delivery depth, and the success of IM agent penetration. Device dynamic pressure measurements were also recorded. RESULTS: ) had the greatest influence on injectate muscle penetration. CONCLUSION: auto-injector achieves delivery depth that ensures intramuscular delivery in both men and women, regardless of BMI. Consistent with other reported data, STMD is greater in women. FUNDING: Crossject.