Differential spectral responsivity describes the spectral responsivity of a solar cell to a low-intensity signal measured under high bias light intensity. As solar cells are used under the high irradiance of the Sun, spectral responsivity of a solar cell has to be measured under similar conditions. Measuring the spectral responsivity of a solar cell under high bias light requires a differential spectral responsivity measurement method. During this work a differential spectral responsivity measurement setup for solar cells was developed. In addition to the instruments measuring the photocurrent of the device under test, the setup comprises two different light sources, a bias light source and a pulsed multichannel LED light source. During this work, the measurement setup was characterized against the spectral irradiance scale of Aalto University. After the characterization, the measurement setup was validated using a calibrated reference solar cell and a custom manufactured reference minimodule. The obtained discrete measurement points were analyzed with Matlab-software. The average difference between the measured and the calibration certificate data was 8.58 per cent. According to the uncertainty analysis, the expanded uncertainty (k=2) of the measurement setup was 11 per cent. To reduce the measurement uncertainty, several methods to improve the setup are suggested at the end of this thesis.