屏蔽电缆
电磁屏蔽
材料科学
核工程
声学
物理
电气工程
工程类
复合材料
作者
P. Koss,Jens-Uwe Voigt,Ronja Rasser,A. Schnabel
出处
期刊:Materials
[MDPI AG]
日期:2024-11-30
卷期号:17 (23): 5877-5877
摘要
A demagnetization study was conducted on a magnetically shielded room (MSR) at Fraunhofer IPM, designed for applications such as magnetoencephalography (MEG) and material testing. With a composite of two layers of mu-metal and an intermediate aluminum layer, the MSR must provide a residual field under 5 nT for the successful operation of optically pumped magnetometers (OPMs). The degaussing process, employing six individual coils, reached the necessary residual magnetic field within the central 1 m3 volume in under four minutes. Due to the low-frequency shielding factor of 100, the obtained average residual field is shown to be limited by environmental residual field changes after degaussing and not by the degaussing procedure.