普鲁士蓝
抗坏血酸
检出限
化学
钴
电化学气体传感器
电化学
无机化学
核化学
尿酸
吸附
色谱法
电极
有机化学
生物化学
物理化学
食品科学
作者
Yanxue Xu,Yunting Qin,Xilan Gao,Jing Li,Dan Xiao
标识
DOI:10.1002/celc.202300134
摘要
Abstract A Prussian blue analogue with cobalt (Na 2‐ x Co[Fe(CN) 6 ] 1‐ y , PB‐Co ) , which is often used as sodium‐ion battery material, was found to be an excellent sensing material toward some biomolecules by calcination. PB‐Co calcined at 750 °C for 1 h (PB‐Co 750‐1 ) has the improved defects to facilitate the adsorption and the catalytic oxidation of ascorbic acid (AA), dopamine (DA) and uric acid (UA). PB‐Co 750‐1 modified on screen printed carbon electrode (SPE) for the ultrasensitively simultaneous determination of AA, DA and UA demonstrates the wide linear range, the low limit of detection (LOD) and the good reproducibility. In addition, the redox of the analytes was confirmed as the diffusion‐controlled process via the kinetic study. The recoveries of AA, DA and UA in real samples were found between 91 % to 114 %. AA and UA in tablets, serums and urines could be determined by PB‐Co 750‐1 /SPE with the satisfactory results.