Continuous and accurate monitoring of volumetric blood flow and other hemodynamic parameters is crucial in various clinical settings. This paper presents a novel wearable ultrasound system based on a cross-shaped probe designed for vector Doppler blood flow monitoring. The probe features a linear-array (LA) transducer for both B-mode imaging and Doppler transmission, complemented by two orthogonal phased-array (PA) segments for multi-angle Doppler signal reception. By capturing Doppler data from two distinct angles, the system compensates for beam-flow misalignment, thereby reducing the angle dependency characteristic of conventional Doppler techniques. Additionally, LA simultaneously acquires cross-sectional images of the target vessel, enabling direct measurement of its cross-sectional area (CSA) for volumetric flow calculations.In vitroexperiments were conducted using a tissue-mimicking phantom and a programmable pump configured to generate pulsatile flow with various stroke volumes. Across this flow range, the proposed approach achieved over 95 % accuracy in volumetric estimations, with only minor deviations. Further tests involving rotational probe angle misalignment demonstrated that measurement fidelity was preserved, underscoring the potential of the system for wearable applications where probe orientation may vary. These results suggest that the cross-shaped probe and its associated vector Doppler methodology can provide a feasible solution for non-invasive, continuous, and wearable hemodynamic monitoring in diverse clinical and ambulatory contexts.