像素
波束赋形
图像质量
计算机科学
扫描线
人工智能
计算机视觉
超声波
工件(错误)
声学
图像(数学)
电信
灰度
物理
作者
Yuhwa Lee,Woo Youl Lee,Chae-Eun Lim,Jin Ho Chang,Tai‐Kyong Song,Yangmo Yoo
出处
期刊:IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control
[Institute of Electrical and Electronics Engineers]
日期:2012-03-01
卷期号:59 (3): 572-582
被引量:16
标识
DOI:10.1109/tuffc.2012.2231
摘要
In this paper, a new compounded direct pixel beamforming (CDPB) method is presented to remove blurring artifacts introduced by ultrasound scan conversion. In CDPB, receive focusing is directly performed on each display pixel in Cartesian coordinates using the raw RF data from adjacent transmit firings so that artifacts from the scan conversion can be removed. In addition, the energy variations resulting from the distance between the transmit scanline and display pixel are compensated by utilizing the gain factor obtained from the ultrasound beam pattern. The proposed CDPB method was evaluated using simulation and in vivo liver data acquired by a commercial ultrasound machine equipped with a research package. The experimental results showed that the proposed CDPB method improved the information entropy contrast (IEC) by 23.6% compared with the conventional scan conversion method and it reduced the blocking artifacts factor (BAF) by 16.4% over the direct pixel-based focusing method. These results indicate the proposed new direct pixel beamforming method could be used to enhance image quality in medical ultrasound imaging.
科研通智能强力驱动
Strongly Powered by AbleSci AI