Several properties of protons make them desirable for radiation therapy. Clinical beam delivery of protons includes proton beams of various energies to deliver a spread-out Bragg peak. Protons are typically produced either in a cyclotron, which generates a monoenergetic continuous beam, or in a synchrotron, in which the protons are delivered in beam spills, but can be delivered at varying energies. A clinical proton beam can be generated either by passive scattering, including the use of a range modulator and second scatterer, or by spot scanning. Regardless of whether passive scattering or spot scanning is used, proton-beam treatment planning is designed to take advantage of the high-dose Bragg peak and steep falloff distal to the Bragg peak. Managing uncertainties is more important for proton therapy than it is for photon therapy, and it is necessary to incorporate knowledge of uncertainties in the design, optimization, and evaluation of proton treatment plans.