闪烁体
剂量计
信号(编程语言)
探测器
闪烁
物理
光学
光子
粒子探测器
辐射
计算机科学
程序设计语言
作者
M A Clift,Peter Johnston,D.V. Webb
标识
DOI:10.1088/0031-9155/47/8/313
摘要
The output signal of an organic scintillator probe consists of a scintillation signal and Cerenkov and fluorescence radiation (CFR) signal when the probe is exposed to a mega-voltage photon or electron beam. The CFR signal is usually unwanted because it comes from both the scintillator and light guide and so it is not proportional to the absorbed dose in the scintillator. A new organic scintillator detector system has been constructed for absorbed dose measurement in pulsed mega-voltage electron and photon beams that are commonly used in radiotherapy treatment, eliminating most of the CFR signal. The new detector system uses a long decay constant BC-444G (Bicron, Newbury, OH, USA) scintillator which gives a signal that can be time resolved from the prompt CFR signal so that the measured contribution of prompt signal is negligible. The response of the new scintillator detector system was compared with the measurements from a plastic scintillator detector that were corrected for the signal contribution from the CFR, and to appropriately corrected ion chamber measurements showing agreement in the 16 MeV electron beam used.
科研通智能强力驱动
Strongly Powered by AbleSci AI