微流控
检测点注意事项
唾液
分析物
纳米技术
材料科学
生物医学工程
色谱法
化学
医学
病理
生物化学
作者
Shixian Liu,Yuting Hou,Zirui Li,Chenyu Yang,Guozhen Liu
出处
期刊:ACS Sensors
[American Chemical Society]
日期:2023-09-05
卷期号:8 (9): 3520-3529
被引量:18
标识
DOI:10.1021/acssensors.3c01093
摘要
A fully integrated device for salivary detection with a sample-in-answer-out fashion is critical for noninvasive point-of-care testing (POCT), especially for the screening of contagious disease infection. Microfluidic paper-based analytical devices (μPADs) have demonstrated their huge potential in POCT due to their low cost and easy adaptation with other components. This study developed a generic POCT platform by integrating a centrifugal microfluidic disc with μPADs to realize sample-to-answer salivary diagnostics. Specifically, a custom centrifugal microfluidic disc integrated with μPADs is fabricated, which demonstrated a high efficiency in saliva treatment. To demonstrate the capability of the integrated device for salivary analysis, the SARS-CoV-2 Nucleocapsid (N) protein, a reliable biomarker for SARS-CoV-2 acute infection, is used as the model analyte. By the chemical treatment of the μPAD surface, and by optimizing the protein immobilization conditions, the on-disc μPADs were able to detect the SARS-CoV-2 N protein down to 10 pg mL-1 with a dynamic range of 10-1000 pg mL-1 and an assay time of 8 min. The integrated device was successfully used for the quantification of the N protein of pseudovirus in saliva with high specificity and demonstrated a comparable performance to the commercial paper lateral flow assay test strips.