光学相干层析成像
医学
脉络膜
眼科
视网膜
视网膜色素上皮
断层摄影术
核医学
视网膜
光学
放射科
物理
作者
Daniel SW Ting,Chui Ming Gemmy Cheung,Laurence Shen Lim,Ian YS Yeo
摘要
Swept source optical coherence tomography (SS-OCT, Topcon Medical System, Japan) is known to have longer wavelength than spectral domain OCT (SD-OCT, Spectralis, Heidelberg Engineering, Germany), allowing a deeper penetration into retina and choroidal layers. This objective of this study was to compare the visibility of retinal and choroidal features in polypoidal choroidal vasculopathy (PCV) using SS-OCT and SD-OCT.This study employs prospective comparative observational case series in Singapore National Eye Center.There were 20 eyes (20 patients) with PCV confirmed with indocyanine green angiogram.Six pre-specified OCT parameters (presence of polyps, sharp pigment epithelial detachment [PED] peak, notched PED and visibility of full maximum height of PED, inner segment/outer segment [IS/OS] line and choroid-scleral interface [CSI]) were graded using SS-OCT and SD-OCT.The Kappa statistics between the two imaging modalities were calculated.Both SS-OCT and SD-OCT were able to detect polypoidal lesions in the majority of eyes (90% and 85%, respectively). However, SS-OCT had better detection for CSI and IS/OS lines (CSI: 80% vs 45%, P = 0.05; IS/OS line: 65% vs 45%, P = 0.34). For sharp PED peak, notched PED, ability to visualize the full PED height and retinal pigment epithelial line, both OCT machines were able to detect in ≥80% of the eyes.In conclusion, SS-OCT and SD-OCT appeared to be similarly effective at detecting most features associated with PCV. However, SS-OCT is more superior in detecting the CSI.