镎
锕系元素
X射线吸收光谱法
扩展X射线吸收精细结构
化学
氧烷
吸收光谱法
光谱学
氧化还原
超铀元素
分析化学(期刊)
吸收(声学)
离子
无机化学
镅
钚
X射线吸收精细结构
放射性废物
放射化学
乏核燃料
物理化学
能斯特方程
铯
作者
Richard Husar,Thomas Dumas,Michel L. Schlegel,Daniel Schlegel,Dominique Guillaumont,Pier‐Lorenzo Solari,Philippe Moisy
标识
DOI:10.1107/s1600577521011115
摘要
A spectroelectrochemical setup has been developed to investigate radioactive elements in small volumes (0.7 to 2 ml) under oxidation–reduction (redox) controlled conditions by X-ray absorption spectroscopy (XAS). The cell design is presented together with in situ XAS measurements performed during neptunium redox reactions. Cycling experiments on the NpO 2 2+ /NpO 2 + redox couple were applied to qualify the cell electrodynamics using XANES measurements and its ability to probe modifications in the neptunyl hydration shell in a 1 mol l −1 HNO 3 solution. The XAS results are in agreement with previous structural studies and the NpO 2 2+ /NpO 2 + standard potential, determined using Nernst methods, is consistent with measurements based on other techniques. Subsequently, the NpO 2 + , NpO 2 2+ and Np 4+ ion structures in solution were stabilized and measured using EXAFS. The resulting fit parameters are again compared with other results from the literature and with theoretical models in order to evaluate how this spectroelectrochemistry experiment succeeds or fails to stabilize the oxidation states of actinides. The experiment succeeded in: (i) implementing a robust and safe XAS device to investigate unstable radioactive species, (ii) evaluate in a reproducible manner the NpO 2 2+ /NpO 2 + standard potential under dilute conditions and (iii) clarify mechanistic aspects of the actinyl hydration sphere in solution. In contrast, a detailed comparison of EXAFS fit parameters shows that this method is less appropriate than the majority of the previously reported chemical methods for the stabilization of the Np 4+ ion.
科研通智能强力驱动
Strongly Powered by AbleSci AI