堵塞
微通道
导电体
收缩
材料科学
电阻抗
微流控
流量(数学)
可用的
粘度
生物医学工程
纳米技术
计算机科学
复合材料
电气工程
机械
工程类
物理
内分泌学
万维网
考古
历史
医学
作者
Junwen Zhu,Yongxiang Feng,Huichao Chai,Fei Liang,Zhen Cheng,Wenhui Wang
出处
期刊:Lab on a Chip
[Royal Society of Chemistry]
日期:2023-01-01
卷期号:23 (11): 2531-2539
被引量:11
摘要
As a label-free and high-throughput single cell analysis platform, impedance flow cytometry (IFC) suffers from clogging caused by a narrow microchannel as mechanical constriction (MC). Current sheath constriction (SC) solutions lack systematic evaluation of the performance and proper guidelines for the sheath fluid. Herein, we hypothesize that the viscosity of the non-conductive liquid is the key to the performance of SC, and propose to employ non-conductive viscous sheath flow in SC to unlock the tradeoff between sensitivity and throughput, while ensuring measurement accuracy. By placing MC and SC in series in the same microfluidic chip, we established an evaluation platform to prove the hypothesis. Through modeling analysis and experiments, we confirmed the accuracy (error < 1.60% ± 4.71%) of SC w.r.t. MC, and demonstrated that viscous non-conductive PEG solution achieved an improved sensitivity (7.92×) and signal-to-noise ratio (1.42×) in impedance measurement, with the accuracy maintained and free of clogging. Viscous SC IFC also shows satisfactory ability to distinguish different types of cancer cells and different subtypes of human breast cancer cells. It is envisioned that viscous SC IFC paves the way for IFC to be really usable in practice with clogging-free, accurate, and sensitive performance.