对映体
色谱法
生物分析
化学
药代动力学
立体选择性
分析化学(期刊)
立体化学
药理学
有机化学
医学
催化作用
作者
Prachi Raikar,Bannimath Gurupadayya,Surulivel Rajan,Sairam Koganti,Peddaguravagari Mounika
出处
期刊:Bioanalysis
[Future Science Ltd]
日期:2022-04-01
卷期号:14 (8): 479-489
标识
DOI:10.4155/bio-2022-0008
摘要
Aim: Ultra-fast LC was used to establish a new bioanalytical method for enantiomeric separation of oxomemazine. Methods: The proposed study was carried out using the ultra-fast LC technique with an amylose chiral column. The bioanalytical approach was used in rabbit plasma following US FDA regulations and then extended to oxomemazine enantiomeric separation using metronidazole as the internal standard. Results: The retention times of (R)-oxomemazine, (S)-oxomemazine and the internal standard were found to be 9.511, 10.712 and 6.503 min, respectively. Within-run and between-run precision (percent relative standard deviation) was found to be in the range of 0.018-0.102% for (R)-oxomemazine and 0.028-0.675% for (S)-oxomemazine, whereas accuracy (%) was found to be in the range of 95.971-99.720% for (R)-oxomemazine and 97.199-103.921% for (S)-oxomemazine. Conclusion: The findings revealed that stereospecific distribution of oxomemazine enantiomers does not change significantly.
科研通智能强力驱动