A Serial Dilution Microfluidic Device for Cytotoxicity Assays
作者
Adrian T. O’Neill,Nancy Ann Monteiro-Riviere,Glenn M. Walker
出处
期刊:Conference proceedings [Institute of Electrical and Electronics Engineers] 日期:2006-08-01被引量:1
标识
DOI:10.1109/iembs.2006.4398036
摘要
A novel microfluidic device is presented which creates a linear serial dilution of two input fluid streams. This platform facilitates higher productivity as a component of a high throughput cytotoxicity testing strategy. A modeling solution is presented to create custom linear dilution schemes. The featured device creates a serial dilution of two solutions in the range of 1:9 through 9:1 across nine discrete dilutions. It has been validated to create a highly linear progression of dilutions with an R2 value of 0.9993. The device functions equivalently over a wide range of flow rates. The standard deviation of dilution values averages 0.76% over six flow rates spanning 0.5 to 16 µl min-1.