Objective:To evaluate the longitudinal flexibility,the radial force,thrombogenecity and radio-opacity of domestic intracranial balloon expandable stent.Methods:Six dogs were divided into three groups at random,stent was implanted into the distal branches of external carotid artery through femoral artery approach by intervention technique.Angiography were performed in 10 minutes (n=12),1 week (n=4),4 weeks (n=4) and 24 weeks (n=4) after stent implantation procedure,the intra-luminal diameter of the target vessel was measured on angiogram accordingly.Results:The domestic stents (n=12) were successfully implanted into the target vessels.Stent dislodgement and balloon rupture were not observed.The stents could be delivered easily to reach the target arteries even crossing through tortuous vessels.and could be expanded to a scheduled diameter after deployment (there were significant difference between the diameters of target vessels before and 10 minutes after implantation,P0.05).No arterial dissection,perforation,ruptures and acute intra-stent thrombosis were found at 10 minutes after stent implantation.An acute arterial occlusion caused by vascular spasm was observed in a small branch distal to the target artery;All target vessels were patent with no stenosis,no stent displacement and no recoils (degree of post-procedural stenosis was less than 50%) at 1 week,4 weeks and 24 weeks after implantation;The stent could be easily visualized under fluoroscopy.Conclusion:The domestic stent has following characteristics: easy delivery,favorable longitudinal flexibility even crossing through tortuous vessels,strong radial force,lower thrombogeneicity and good radio-opacity.