医学
介入放射学
辐射暴露
剂量计
核医学
放射科
剂量学
作者
Thiago Viana Miranda Lima,Tomás Reyes del Castillo,Mirjam Heinrich,Stefanie Zihlmann,Rubén López Benítez,Justus E. Roos
标识
DOI:10.1088/1361-6498/ac4e13
摘要
To evaluate the usefulness of commercially available scatter reduction drapes in mitigating staff exposure in interventional radiology and the potential harmful effects of drape malpositioning in terms of exposure levels to both patients and staff. An anthropomorphic phantom was irradiated on an angiography device under three scenarios: no drape and correct and incorrect drape positioning. Different levels of incorrect drape positioning relative to the field-of-view (FOV) were evaluated: slight, mild and severe. Real-time dosimeter systems (positioned on the operator's eye, chest and thyroid) were used to evaluate accumulative doses and dose rates. Different obstruction levels were evaluated and compared to the observer's perception. Additionally, patient exposure was evaluated for all scenarios using a dose area product (DAP). Up to a mild obstruction, by using the drape a dose reduction of up to 86% was obtained while a severe obstruction produced a 1000% increase in exposure, respectively for all dosimeter positions compared to the use of no drape. A similar order of magnitude was observed for patient exposure. Good agreement was obtained for the observer perception of the FOV obstruction up to 25% of the FOV; for larger obstructions, an overestimate of the obstruction was observed. Patient lead drapes can reduce staff doses in interventional radiology procedures even when mildly malpositioned and obscuring the FOV. Special attention to protective drape positioning is necessary, since the severe obstruction of the FOV results in a large increase in both operator and patient exposure.
科研通智能强力驱动
Strongly Powered by AbleSci AI