The immunity and survivability in a strong electromagnetic environment make the Faraday Rotation (FR) method irreplaceable for measuring pulsed magnetic fields. We present a dual-channel FR method that achieves more stable and precise measurement and more applicable magnetic field types than the conventional FR method. By splitting the modulated light into two components, the ratio of the two components is transformed into a tangent function with two analyzers at a relative 45°. The use of terbium gallium garnet, which has a significant magneto-optical effect, as a magneto-optical medium enables higher sensitivity for measurement. We have successfully measured the variation in magnetic flux density of the capacitively driven pulsed magnetic field.