机器人
接触力
计算机科学
材料科学
物理
人工智能
经典力学
作者
Majid Roshanfar,Pedram Fekri,Robert Nguyen,Changyan He,Paul H. Kang,James M. Drake,Eric Diller,Thomas Looi
出处
期刊:Proceedings
[Institute of Electrical and Electronics Engineers]
日期:2025-05-19
卷期号:: 844-850
被引量:6
标识
DOI:10.1109/icra55743.2025.11128662
摘要
Knowledge of the tip contact force in continuum robots, which are often used as medical instruments, is critical for clinical applications. It enhances the interventionalist's decision-making, navigation efficiency, and procedural safety. However, accurately determining the tip contact force in conventionally sized instruments remains challenging. This study introduces a learning-based method for estimating the external contact force at the tip of a continuum robot. By leveraging curvature and bending angle data from a multi-core fiber equipped with fiber Bragg gratings (FBGs) embedded inside the Nitinol tube, the method maps these inputs to the corresponding tip force in 3D. Experiments conducted on an FBG-embedded Nitinol rod validate the feasibility of the proposed method, yielding Mean Squared Error (MSE), Mean Absolute Error (MAE), and Root Mean Squared Error (RMSE) values of 20.9$\left(m N^{2}\right), 2.7(m N)$, and$4.6(m N)$, respectively, which represent a 26 % improvement compared to the learning-based vision methodology.