Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities.

溶血酶- 扫描仪 成像体模 图像分辨率 核医学 材料科学 物理 视野 光学 探测器 闪烁体 医学
作者
Suleman Surti,A. Kuhn,Matthew E. Werner,Amy E. Perkins,Jeffrey Kolthammer,Joel S. Karp
出处
期刊:PubMed 卷期号:48 (3): 471-80 被引量:150
链接
标识
摘要

Results from a new PET/CT scanner using lutetium-yttrium oxyorthosilicate (LYSO) crystals for the PET component are presented. This scanner, which operates in a fully 3-dimensional mode, has a diameter of 90 cm and an axial field of view of 18 cm. It uses 4 x 4 x 22 mm(3) LYSO crystals arranged in a pixelated Anger-logic detector design. This scanner was designed to perform as a high-performance conventional PET scanner as well as provide good timing resolution to operate as a time-of-flight (TOF) PET scanner.Performance measurements on the scanner were made using the National Electrical Manufacturers Association (NEMA) NU2-2001 procedures to benchmark its conventional imaging capabilities. The scatter fraction and noise equivalent count (NEC) measurements with the NEMA cylinder (20-cm diameter) were repeated for 2 larger cylinders (27-cm and 35-cm diameter), which better represent average and heavy patients. New measurements were designed to characterize its intrinsic timing resolution capability, which defines its TOF performance. Additional measurements to study the impact of pulse pileup at high counting rates on timing, as well as energy and spatial, resolution were also performed. Finally, to characterize the effect of TOF reconstruction on lesion contrast and noise, the standard NEMA/International Electrotechnical Commission torso phantom as well as a large 35-cm-diameter phantom with both hot and cold spheres were imaged for varying scan times.The transverse and axial resolution near the center is 4.8 mm. The absolute sensitivity of this scanner measured with a 70-cm-long line source is 6.6 cps/kBq, whereas scatter fraction is 27% measured with a 70-cm-long line source in a 20-cm-diameter cylinder. For the same line source cylinder, the peak NEC rate is measured to be 125 kcps at an activity concentration of 17.4 kBq/mL (0.47 microCi/mL). The 2 larger cylinders showed a decrease in the peak NEC due to increased attenuation, scatter, and random coincidences, and the peak occurs at lower activity concentrations. The system coincidence timing resolution was measured to be 585 ps. The timing resolution changes as a function of the singles rate due to pulse pileup and could impact TOF image reconstruction. Image-quality measurements with the torso phantom show that very high quality images can be obtained with short scan times (1-2 min per bed position). However, the benefit of TOF is more apparent with the large 35-cm-diameter phantom, where small spheres are detectable only with TOF information for short scan times.The Gemini TF whole-body scanner represents the first commercially available fully 3-dimensional PET scanner that achieves TOF capability as well as conventional imaging capabilities. The timing resolution is also stable over a long duration, indicating the practicality of this device. Excellent image quality is achieved for whole-body studies in 10-30 min, depending on patient size. The most significant improvement with TOF is seen for the heaviest patients.

科研通智能强力驱动
Strongly Powered by AbleSci AI
科研通是完全免费的文献互助平台,具备全网最快的应助速度,最高的求助完成率。 对每一个文献求助,科研通都将尽心尽力,给求助人一个满意的交代。
实时播报
朴实的面包完成签到,获得积分20
刚刚
DDDD发布了新的文献求助10
刚刚
1秒前
HUYAOWEI发布了新的文献求助10
1秒前
2秒前
Li完成签到 ,获得积分10
2秒前
2秒前
冰块儿发布了新的文献求助10
3秒前
3秒前
GRX1110发布了新的文献求助10
3秒前
3秒前
桐桐应助隐形小熊猫采纳,获得10
4秒前
4秒前
5秒前
lelecha发布了新的文献求助10
6秒前
6秒前
I北草蜥完成签到,获得积分10
6秒前
botion发布了新的文献求助10
7秒前
小猫咪完成签到,获得积分10
7秒前
7秒前
居崽完成签到 ,获得积分10
7秒前
boom发布了新的文献求助10
7秒前
8秒前
打打应助大梦采纳,获得10
8秒前
科研通AI6应助fnuew采纳,获得10
8秒前
8秒前
量子星尘发布了新的文献求助10
9秒前
李爱国应助Dream采纳,获得10
9秒前
9秒前
9秒前
zhl发布了新的文献求助10
10秒前
CodeCraft应助lixm采纳,获得10
10秒前
邱天完成签到,获得积分10
10秒前
11秒前
11秒前
科研通AI6应助botion采纳,获得10
12秒前
13秒前
13秒前
yqq38发布了新的文献求助10
14秒前
草帽发布了新的文献求助10
14秒前
高分求助中
(应助此贴封号)【重要!!请各用户(尤其是新用户)详细阅读】【科研通的精品贴汇总】 10000
Encyclopedia of Reproduction Third Edition 3000
《药学类医疗服务价格项目立项指南(征求意见稿)》 1000
花の香りの秘密―遺伝子情報から機能性まで 800
1st Edition Sports Rehabilitation and Training Multidisciplinary Perspectives By Richard Moss, Adam Gledhill 600
nephSAP® Nephrology Self-Assessment Program - Hypertension The American Society of Nephrology 500
Digital and Social Media Marketing 500
热门求助领域 (近24小时)
化学 材料科学 生物 医学 工程类 计算机科学 有机化学 物理 生物化学 纳米技术 复合材料 内科学 化学工程 人工智能 催化作用 遗传学 数学 基因 量子力学 物理化学
热门帖子
关注 科研通微信公众号,转发送积分 5626305
求助须知:如何正确求助?哪些是违规求助? 4712106
关于积分的说明 14958305
捐赠科研通 4781298
什么是DOI,文献DOI怎么找? 2554214
邀请新用户注册赠送积分活动 1515957
关于科研通互助平台的介绍 1476289