膝关节
十字韧带
拉赫曼试验
股骨
后交叉韧带
前交叉韧带
流离失所(心理学)
医学
胫骨
口腔正畸科
骨科手术
外科
前交叉韧带重建术
心理学
心理治疗师
作者
F Kryszohn,Marilyn Luber,Ellen F. Binder,P Schaff
出处
期刊:Biomedizinische Technik
[De Gruyter]
日期:1999-01-01
卷期号:44 (7-8): 202-205
标识
DOI:10.1515/bmte.1999.44.7-8.202
摘要
The KT2000 knee ligament arthrometer manufactured by the MEDmetric Corporation is used in the field of orthopedics to establish quantitatively the size of the anterior and posterior draw phenomena of the knee joint. The basis for the measurement is the Lachman test in which a force is applied to the tibia relative to the femur. The resulting displacement is limited by the cruciate ligaments. The measurements obtained for a potentially injured knee are compared with those of the healthy knee and empirical threshold values, and serve as a basis for deciding whether to operate on patients with cruciate ligament insufficiency. The subject of the present study is a design shortcoming of the KT2000 that may falsify the measured results when the device is not correctly employed. A simple mechanical modification of the KT2000 can reduce this error. In the present study, the error is quantified and we show how the modification can keep it within tolerable limits.
科研通智能强力驱动
Strongly Powered by AbleSci AI