摘要
Decreased cardiac function can have a negative impact on other organs. The left ventricular pressure-volume relationship is considered to be a valid method for evaluating cardiac function. Real-time monitoring of cardiac function is important for drug evaluation. Under closed-chest conditions, the miniature transducer, which is an important component of the pressure-volume catheter, enters the left ventricle of the rat through the right carotid artery. The device visualizes the changes in cardiac function during the experiment in the form of a pressure-volume loop. The actual volume of the ventricle is calculated by altering the conductivity of the blood by injecting 50 µL of a 20% sodium chloride solution into the rat's left jugular vein. The actual volume of the rat's ventricular cavity is calculated by measuring the conductivity of the blood in a known volume using a pressure-volume conductance catheter. This protocol allows for continuous observation of the effects of drugs on the heart and will promote the rationale for the use of specialty ethnic drugs in cardiovascular disease.