医学
单室膝关节置换术
还原(数学)
位错
外翻
骨关节炎
撑杆
方位(导航)
外科
口腔正畸科
地理
地图学
复合材料
工程类
病理
数学
替代医学
几何学
材料科学
机械工程
作者
Kohei Kawaguchi,Hiroshi Inui,Shuji Taketomi,Ryota Yamagami,Kenichi Kono,Keiu Nakazato,Manabu Kawata,Sakae Tanaka
出处
期刊:Knee
[Elsevier BV]
日期:2018-12-15
卷期号:26 (1): 267-272
被引量:14
标识
DOI:10.1016/j.knee.2018.10.015
摘要
Meniscal bearing dislocation while rolling over in sleep has never been reported in Oxford unicompartmental knee arthroplasty (UKA). This study reports two cases of meniscal bearing dislocation into the intercondylar ridge while rolling over in sleep. In the case of one patient, closed reduction of the bearing was performed, and the use of a knee brace was effective in preventing re-dislocation. In the second patient, closed reduction was possible; however, bearing dislocation was repeated. Therefore, revision surgery was performed by replacing the tibial component and using a thicker bearing. The common features in dislocation during rolling over while sleeping in both cases were dislocation into the intercondylar ridge, the combination of small femur and AA-size tibia components, and osteonecrosis. As determined by intraoperative testing, valgus position of the knee while rolling over in sleep could induce bearing dislocation into the intercondylar ridge. LEVEL OF EVIDENCE: Retrospective case series, Level IV.