成像体模
图像分辨率
光子计数
探测器
图像质量
材料科学
光学
噪音(视频)
闪烁体
核医学
医学物理学
物理
计算机科学
人工智能
医学
图像(数学)
作者
Xiaohui Zhan,Ruoqiao Zhang,Xiaofeng Niu,I.A. Hein,Brent Budden,Shuoxing Wu,Nicolay Markov,C. Clarke,Y. Qiang,Hiroki Taguchi,Keiichi Nomura,Yoshihisa Muramatsu,Yu Zhou,Tatsushi Kobayashi,Richard Thompson,Hiroaki Miyazaki,Hiroaki Nakai
标识
DOI:10.1088/1361-6560/acebb3
摘要
Objective. Photon counting CT (PCCT) has been a research focus in the last two decades. Recent studies and advancements have demonstrated that systems using semiconductor-based photon counting detectors (PCDs) have the potential to provide better contrast, noise and spatial resolution performance compared to conventional scintillator-based systems. With multi-energy threshold detection, PCD can simultaneously provide the photon energy measurement and enable material decomposition for spectral imaging. In this work, we report a performance evaluation of our first CdZnTe-based prototype full-size PCCT system through various phantom imaging studies.Approach.This prototype system supports a 500 mm scan field-of-view and 10 mmz-coverage at isocenter. Phantom scans were acquired using 120 kVp from 50 to 400 mAs to assess the imaging performance on: CT number accuracy, uniformity, noise, spatial resolution, material differentiation and quantification.Main results.Both qualitative and quantitative evaluations show that PCCT, under the tested conditions, has superior imaging performance with lower noise and improved spatial resolution compared to conventional energy integrating detector (EID)-CT. Using projection domain material decomposition approach with multiple energy bin measurements, PCCT virtual monoenergetic images have lower noise, and good accuracy in quantifying iodine and calcium concentrations. These results lead to increased contrast-to-noise ratio (CNR) for both high and low contrast study objects compared to EID-CT at matched dose and spatial resolution. PCCT can also generate super-high resolution images using much smaller detector pixel size than EID-CT and greatly improve image spatial resolution.Significance.Improved spatial resolution and quantification accuracy with reduced image noise of the PCCT images can potentially lead to better diagnosis at reduced radiation dose compared to conventional EID-CT. Increased CNR achieved by PCCT suggests potential reduction in iodine contrast media load, resulting in better patient safety and reduced cost.
科研通智能强力驱动
Strongly Powered by AbleSci AI