Dusan Pisarcik,Jakob Heimer,Monika Farkas,R. Huch,Cornelia Leo
出处
期刊:Ejso
[Elsevier BV]
日期:2025-10-06
卷期号:51 (12): 110495-110495
标识
DOI:10.1016/j.ejso.2025.110495
摘要
We evaluated the magnetic Pintuition marker for stereotactic localization of non-palpable breast lesions and compared deployment techniques in a phantom model. In a retrospective single-center cohort (March 2023-January 2025), marker-to-target distances were recorded immediately after placement and, when available, on the day of surgery; malposition/dislocation used a 10 mm threshold (dislocation also at 5 mm). Across 67 placements in 53 patients, the median distance was 2.8 mm (IQR 0.0-6.7 mm); malposition occurred in 16.7 %. No association with ACR density or approach was detected. Among 33 procedures with follow-up imaging, dislocation was rare (3 %, n = 1, using ≥10 mm threshold and 6.1 %, n = 2, at ≥5 mm). Initial distances and malposition rates declined over time, with no malpositions after July 2024. In phantom experiments, all 15 placements were within 3 mm. Findings indicate precise, stable targeting and support Pintuition as a flexible alternative to traditional localization for breast-conserving surgery.