Inspection of endoscope instrument channels after reprocessing using a prototype borescope

医学 内窥镜 频道(广播) 生物医学工程 外科 计算机科学 计算机网络
作者
Adarsh M. Thaker,Stephen Kim,Alireza Sedarat,Rabindra R. Watson,V. Raman Muthusamy
出处
期刊:Gastrointestinal Endoscopy [Elsevier BV]
卷期号:88 (4): 612-619 被引量:80
标识
DOI:10.1016/j.gie.2018.04.2366
摘要

Background and Aims Visual inspection of the instrument channel has been proposed as a quality assurance step during endoscope reprocessing. However, the nature and severity of findings in a broad array of endoscopes (gastroscopes, colonoscopes, duodenoscopes, and echoendoscopes) after systemic implementation of an inspection protocol remain unknown. In addition, a study using borescope inspection in upper endoscopes and colonoscopes raised concerns about persistent simethicone residue despite full reprocessing. Methods A pilot inspection study using a prototype borescope (SteriCam Inspection Scope; SteriView Inc, San Rafael, Calif) was performed on routinely used endoscopes after high-level disinfection, manual forced-air dry of the instrument channel, and overnight vertical storage. Video recordings 1.5 to 2 minutes in duration were reviewed for visible moisture, debris, discoloration, scratches, channel shredding (scratches that result in strips or filaments of the channel lining protruding into the lumen), and visible evidence of biofilm or simethicone residue. Results A total of 97 inspections of 59 endoscopes were reviewed. The most common finding was scratches, seen in 51 devices (86%). Channel shredding was found in 35 devices (59%). Intrachannel debris was identified in 22 (23%) of the 97 inspections. No moisture was seen (0%) in the 74 inspections performed after forced-air dry and overnight vertical storage compared with moisture in 5 of 18 inspections (28%) performed after storage alone. No visual evidence of biofilm or simethicone residue was discovered despite its frequent use in our unit. Conclusion Internal defects of the instrument channel appear to occur frequently. Manual forced-air drying of the channel appears to be highly effective in eliminating moisture compared with overnight hang drying alone. Video inspection of the endoscope channel may be useful to audit reprocessing performance and to identify damaged endoscopes. Visual inspection of the instrument channel has been proposed as a quality assurance step during endoscope reprocessing. However, the nature and severity of findings in a broad array of endoscopes (gastroscopes, colonoscopes, duodenoscopes, and echoendoscopes) after systemic implementation of an inspection protocol remain unknown. In addition, a study using borescope inspection in upper endoscopes and colonoscopes raised concerns about persistent simethicone residue despite full reprocessing. A pilot inspection study using a prototype borescope (SteriCam Inspection Scope; SteriView Inc, San Rafael, Calif) was performed on routinely used endoscopes after high-level disinfection, manual forced-air dry of the instrument channel, and overnight vertical storage. Video recordings 1.5 to 2 minutes in duration were reviewed for visible moisture, debris, discoloration, scratches, channel shredding (scratches that result in strips or filaments of the channel lining protruding into the lumen), and visible evidence of biofilm or simethicone residue. A total of 97 inspections of 59 endoscopes were reviewed. The most common finding was scratches, seen in 51 devices (86%). Channel shredding was found in 35 devices (59%). Intrachannel debris was identified in 22 (23%) of the 97 inspections. No moisture was seen (0%) in the 74 inspections performed after forced-air dry and overnight vertical storage compared with moisture in 5 of 18 inspections (28%) performed after storage alone. No visual evidence of biofilm or simethicone residue was discovered despite its frequent use in our unit. Internal defects of the instrument channel appear to occur frequently. Manual forced-air drying of the channel appears to be highly effective in eliminating moisture compared with overnight hang drying alone. Video inspection of the endoscope channel may be useful to audit reprocessing performance and to identify damaged endoscopes.
最长约 10秒,即可获得该文献文件

科研通智能强力驱动
Strongly Powered by AbleSci AI
科研通是完全免费的文献互助平台,具备全网最快的应助速度,最高的求助完成率。 对每一个文献求助,科研通都将尽心尽力,给求助人一个满意的交代。
实时播报
1秒前
李爱国应助wtbxsjy采纳,获得10
1秒前
小白发布了新的文献求助10
1秒前
2秒前
shawnho完成签到,获得积分10
2秒前
四硼酸钠完成签到,获得积分10
2秒前
3秒前
chengche发布了新的文献求助10
4秒前
拼搏向上发布了新的文献求助10
4秒前
4秒前
大方的依琴完成签到 ,获得积分10
4秒前
虚心香彤发布了新的文献求助10
5秒前
ly发布了新的文献求助10
5秒前
Calista发布了新的文献求助10
7秒前
7秒前
优雅沛文完成签到,获得积分10
9秒前
wtbxsjy发布了新的文献求助10
9秒前
10秒前
C胖胖完成签到,获得积分10
10秒前
罗亚亚发布了新的文献求助10
10秒前
枝瓯发布了新的文献求助10
12秒前
Kiki完成签到,获得积分10
13秒前
彭于晏应助王大伟2023采纳,获得10
13秒前
小二郎应助王大伟2023采纳,获得10
14秒前
烟花应助王大伟2023采纳,获得10
14秒前
思源应助王大伟2023采纳,获得10
14秒前
星辰大海应助王大伟2023采纳,获得10
14秒前
上官若男应助王大伟2023采纳,获得10
14秒前
Owen应助王大伟2023采纳,获得10
14秒前
14秒前
乐乐应助王大伟2023采纳,获得10
14秒前
小蘑菇应助王大伟2023采纳,获得10
14秒前
在水一方应助王大伟2023采纳,获得10
14秒前
朴实天曼发布了新的文献求助10
15秒前
迷路的翼完成签到,获得积分10
15秒前
15秒前
Bellis完成签到 ,获得积分10
16秒前
大锤发布了新的文献求助10
16秒前
77完成签到 ,获得积分10
17秒前
17秒前
高分求助中
Principles of Economics, 11th Edition 10000
Prescott's Microbiology: 2026 Release ISE 10000
University Physics with Modern Physics, 16th edition 10000
(应助此贴封号)【重要!!请各用户(尤其是新用户)详细阅读】【科研通的精品贴汇总】 10000
Environmental Leverage in Times of Climate Crisis: Product Standards, Carbon Border Measures and Preferential Trade Agreements 1000
Interactions of Vowel Quality and Prosody in East Slavic 1000
Erwählung und Berufung bei Paulus: Bedeutung, Entwicklung und Funktion einer Vorstellung in ihrem frühjüdischen und griechisch-römischen Kontext 850
热门求助领域 (近24小时)
化学 材料科学 医学 生物 纳米技术 工程类 有机化学 化学工程 生物化学 计算机科学 内科学 物理 复合材料 催化作用 细胞生物学 无机化学 光电子学 物理化学 电极 基因
热门帖子
关注 科研通微信公众号,转发送积分 7196935
求助须知:如何正确求助?哪些是违规求助? 8832088
关于积分的说明 18646049
捐赠科研通 6835573
什么是DOI,文献DOI怎么找? 3177267
关于科研通互助平台的介绍 2331133
邀请新用户注册赠送积分活动 2151817