Manli Zhao,Qianyu Zhang,Dongfang Li,Chao Tao,Xiaojun Liu
出处
期刊:Optics Express
[Optica Publishing Group]
日期:2023-12-19
卷期号:32 (2): 1501-1501
被引量:2
摘要
In this study, we designed a self-focused ultrasonic transducer made of polyvinylidene fluoride (PVDF). This transducer involves a back-reflector, which is modeled after tapetum lucidum in the eyes of some nocturnal animals. The bionic structure reflects the ultrasound, which passes through the PVDF membrane, back to PVDF and provides a second chance for the PVDF to convert the ultrasound to electric signals. This design increases the amount of ultrasound absorbed by the PVDF, thereby improving the detection sensitivity. Both ultrasonic and photoacoustic (PA) experiments were conduct to characterize the performance of the transducer. The results show that the fabricated transducer has a center frequency of 13.07 MHz, and a bandwidth of 96% at -6 dB. With an acoustic numerical aperture (NA) of 0.64, the transducer provides a lateral resolution of 140µm. Importantly, the bionic design improves the detection sensitivity of the transducer about 30%. Finally, we apply the fabricated transducer to optical-resolution (OR) and acoustic-resolution photoacoustic microscopy (AR-PAM) to achieve multiscale-resolution PA imaging. Imaging of the bamboo leaf and the leaf skeleton demonstrates that the proposed transducer can provide high spatial resolution, better imaging intensity and contrast. Therefore, the proposed transducer design will be useful to enhance the performance of multiscale-resolution PAM.