漫反射光学成像
成像体模
光学
迭代重建
规范化(社会学)
光学层析成像
计算机科学
光学相干层析成像
多光谱图像
断层摄影术
计算机视觉
生物医学工程
物理
医学
人类学
社会学
作者
Jingjing Jiang,Aldo Di Costanzo Mata,Scott Lindner,Chao Zhang,Edoardo Charbon,Martin Wolf,Alexander Kalyanov
摘要
Near infrared optical tomography (NIROT) is an emerging modality that enables imaging the oxygenation of tissue, which is a biomarker of tremendous clinical relevance. Measuring in reflectance is usually required when NIROT is applied in clinical scenarios. Single photon avalanche diode (SPAD) array technology provides a compact solution for time domain (TD) NIROT to gain huge temporal and spatial information. This makes it possible to image complex structures in tissue. The main aim of this paper is to validate the wavelength normalization method for our new TD NIROT experimentally by exposing it to a particularly difficult challenge: the recovery of two inclusions at different depths. The proposed reconstruction algorithm aims to tackle systematic errors and other artifacts with known wavelength-dependent relation. We validated the device and reconstruction method experimentally on a silicone phantom with two inclusions: one at depth of 10 mm and the other at 15 mm. Despite this tough challenge for reflectance NIROT, the system was able to localize both inclusions accurately.
科研通智能强力驱动
Strongly Powered by AbleSci AI