融合
可预测性
传感器融合
拉曼光谱
标准差
保险丝(电气)
化学
分析化学(期刊)
计算机科学
生物系统
数学
人工智能
色谱法
统计
生物
光学
工程类
物理
哲学
语言学
电气工程
作者
Chima Robert,Wendy E. Bain,Cameron R. Craigie,Talia Maree Hicks,M.P.F. Loeffen,Sara J. Fraser‐Miller,Keith Christopher Gordon
出处
期刊:Meat Science
[Elsevier BV]
日期:2022-10-14
卷期号:195: 109005-109005
被引量:17
标识
DOI:10.1016/j.meatsci.2022.109005
摘要
The application of individual spectroscopic techniques for meat analysis has been widely explored. Attempts to fuse data from multiple spectroscopic instruments for meat analysis are still lacking. Comparative assessment of the performance of mid infrared (MIR), near infrared (NIR) and Raman spectroscopy to estimate fatty acid (FA) composition in processed lamb was investigated. The acquired data from these individual techniques were then utilised in estimating similar parameters using a multi-block partial least square data fusion approach. Model performance was assessed with respect to the determination coefficient and ratio of predictive deviation upon cross-validation of the model. The fused data had slight improvements for the prediction of four FA parameters including MUFA, C18:0, C18:1 c9 and C9, t11- CLA), suggesting possible information enhancement with use of multiple instruments. However, MIR offered better predictability (RPD values) across the FA parameters considered. • Fusion of data blocks from three spectroscopic techniques was investigated for fatty acid prediction. • Data fusion yielded good prediction performance for the different fatty acid compositions. • Data fusion yielded improved performance on comparison to individual techniques.
科研通智能强力驱动