电容式微机械超声换能器
材料科学
超声波传感器
制作
传感器
表面微加工
电容感应
蚀刻(微加工)
基质(水族馆)
光电子学
声学
电气工程
复合材料
工程类
图层(电子)
替代医学
病理
地质学
物理
海洋学
医学
作者
Oluwafemi Joel Adelegan,Zachary A. Coutant,Xiao Zhang,Feysel Yalcin Yamaner,Ömer Oralkan
出处
期刊:
[Institute of Electrical and Electronics Engineers]
日期:2019-10-01
卷期号:: 1205-1208
被引量:5
标识
DOI:10.1109/ultsym.2019.8925783
摘要
A two-dimensional (2D) transducer array is an integral part of a three-dimensional (3D) ultrasound imaging system as well as a compact ultrasound system for neurostimulation to steer and focus the beam in a volume. In this paper, a sacrificial etching-based fabrication process for implementing a 16x16-element 2D CMUT array on a glass substrate with through-glass-interconnects is described in detail. Across the fabricated 256 elements of the 2D CMUT array, the mean resonant frequency is measured as 4.76 MHz with a standard deviation of 46.6 kHz. The fabricated 2D CMUT array shows a 100% element yield in fabrication and excellent uniformity in device performance. The process offers the advantages of developing 2D CMUT arrays on glass substrates that do not need to be compatible with anodic bonding.