分析化学(期刊)
溶解
感应耦合等离子体
电感耦合等离子体质谱法
化学
等离子体原子发射光谱
NIST公司
原子发射光谱法
标准溶液
光谱学
质谱法
材料科学
等离子体
色谱法
物理
物理化学
量子力学
计算机科学
自然语言处理
作者
Frank M. Pennebaker,M. Bonner Denton
标识
DOI:10.1366/0003702011952037
摘要
A robust method of analysis that uses multichannel array detector technology is presented for high-precision, accurate measurement of Pt, Pd and Rh in catalytic converters. Catalytic converter samples, National Institute of Standards and Technology (NIST) reference standards, and a fresh converter sample are dissolved and analyzed by inductively coupled plasma atomic emission spectroscopy (ICP-AES). Complete dissolution of sample sizes up to 2 g of NIST reference standards is achieved with the use of high-pressure, high-temperature carius tubes. Carius tube dissolution of a modern catalytic converter sample is found to be incomplete because of the presence of higher concentrations of Ce, Ni, and Zr. For this sample, the addition of FeCl 3 increases the dissolving power of the carius tube for base metals and noble metals, particularly Rh. Although interference problems are encountered in samples bearing high concentrations of Fe, use of a high-resolution ICP-AES system alleviated these problems. Precise quantitation of Pt, Pd, and Rh is accomplished by using the method of standard additions, with high accuracy and a precision of 1–2% RSD (relative standard deviation). This analysis also confirmed the values determined by NIST by using isotope dilution inductively coupled plasma mass spectrometry (ICP-MS) for standard reference materials (SRMs) 2556 and 2557.
科研通智能强力驱动
Strongly Powered by AbleSci AI