触针
重复性
偏转(物理)
准确度和精密度
有限元法
光学
长度测量
观测误差
声学
材料科学
复制
测量仪器
统计分析
数学
测量不确定度
误差分析
实验研究
量具(枪械)
工程类
轮廓仪
计算机科学
作者
Parth Gaba,Yash Kumar,Arnab Haldar,Girija Moona,Jitendra Kumar,Vinod Kumar,Kunal Jangid
出处
期刊:
[Institute of Electrical and Electronics Engineers]
日期:2025-10-31
卷期号:: 1-5
标识
DOI:10.1109/delcon68055.2025.11400289
摘要
This study combines experimental measurements and finite element simulations to quantify the effect of stylus length on CMM (Coordinate Measuring Machine) accuracy. Two standard artifacts, a ceramic sphere and a step gauge, were subjected to dimensional evaluations using styli with stem lengths of 50 mm and 75 mm. The investigation findings exhibit that, as a result of increased deflection and dynamic instability, longer stylus lengths demonstrate higher measurement errors: 5.3 times error increment in sphere diameter and 6.2 times error increment in step gauge length measurements, using 75 mm vs 50 mm stylus length. Additionally, the statistical analysis reveals that the 75 mm stylus display unstable dimensional accuracy with inferior repeatability; 6.5 times repeatability abatement than the 50 mm long stylus. Simulations replicate the experimental deflection trends, demonstrating the effectiveness of FEM-based modeling in guiding optimal stylus selection for precision metrology.
科研通智能强力驱动
Strongly Powered by AbleSci AI